Purpose
To measure the airborne particulate level in the cleanroom and to determine the room cleanliness grade with reference to ISO 14644-1, EU GMP classifications at the time of the test (As-Built, At-Rest, or Operational).
Instrument
Discrete Particle Counter
Procedure
- Verify that all aspects of the cleanroom system which contribute to its operational integrity (air handling, filtration systems, walls, ceilings, floors, etc.) are complete and functioning. The following primary tests shall be completed prior to performing the cleanliness classification tests:
- Airflow test
- Filter leak test
- Room pressurization test
- The Cleanroom has been purged for at least 12 hours before testing commences.
- The rooms are divided into test grids as per minimum sampling locations requirement by ISO 14644-1:2015.
- The counter’s sample rate is set to 1 m3/min and the sampling time set to 1 minute or as required.
- A minimum of 1 sample is taken at each grid, at a height of 1 meter above the floor.
- When an obstruction is encountered, the sample is to be taken at 300 mm above the obstruction.